You are here
Biological Safety Cabinet Class II A2
Features
- Motorized Front Window.
- Time Reserve Function.
- HEPA Filter life and UV Life indicator.
- Automatic air speed adjustable with filter block.
- Work area surrounded by negative pressure, it can ensure maximum safety in work area.
- Audio and visual alarm (Filter replacement, Window over height, abnormal Air flow velocity etc).
- Most accessories are standard. No need to pay more, save your money.
- Remote Control. All functions can be realized with it, making the operation much easier and more convenient.
- Foot Switch. Adjust front window height by foot during experiment, to avoid airflow turbulence caused by arm movement.
- Interlock Function: UV lamp and front window; UV lamp and blower, fluorescent lamp; Blower and front window.
Models
Model |
BSC-1100IIA2-X |
BSC-1300IIA2-X |
BSC-1500IIA2-X |
BSC-1800IIA2-X |
BSC-2000IIA2-X |
Internal Size (W*D*H) |
940* 600*660 mm |
1150*600*660 mm |
1350*600*660mm |
1710*600*660 mm |
1800*600*660mm |
External Size (W*D*H) |
1100*750*2250mm |
1300*750*2250mm |
1500*750*2250mm |
1870*750*2290mm |
1950*800*2305mm |
Tested Opening |
Safety Height 200 mm(8’’) |
||||
Max Opening |
420mm(17’’) |
420mm(17’’) |
400mm(16’’) |
400mm(16’’) |
400mm(16’’) |
Inflow Velocity |
0.53±0.025 m/s |
||||
Downflow Velocity |
0.33±0.025 m/s |
||||
HEPA Filter |
Two, 99.999% efficient at 0.3 um. Filter life indicator. |
||||
Front Window |
Motorized. Two-layer laminated toughened glass ≥ 5mm. Anti UV. |
||||
Noise |
EN12469 ≤ 58 dB / NSF49 ≤ 61 dB |
||||
UV Lamp |
30W*1 |
30W* 1 |
40W* 1 |
40W*2 |
40W* 1 |
UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination. |
|||||
Fluorescent Lamp |
21W*2 |
21W*2 |
28W*2 |
40W*2 |
40W*2 |
Illumination |
≥1000Lux |
||||
Consumption |
760W |
800W |
900W |
900W |
1500W |
Waterproof Sockets |
Two, total load of two sockets: 500W |
||||
Display |
LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow velocity, filter life, humidity and temperature, system working time etc. |
||||
Control System |
Microprocessor |
||||
Airflow System |
70% air recirculation, 30% air exhaust |
||||
Visual and audio alarm |
Filter replacement, window over height, abnormal air flow velocity. |
||||
Material
|
Work Zone: 304 stainless steel |
||||
Main Body: Cold-rolled steel with anti-bacteria powder coating. |
|||||
Work Surface Height |
750mm
|
||||
Caster |
Footmaster caster |
||||
Power Supply |
AC220V±10%, 50/60Hz; 110V±10%, 60Hz |